  Defibrillator lead migration with subacute right ventricular perforation : A case report
 
   
   
     
 
Natchayathipk Kittichamroen
Internal medicine department, Charoenkrung Pracharak Hospital, Department of Medical Services, Bangkok Metropolitan Administration (BMA)

Abstract

Permanent pacemaker (PPM) or automated implantable cardioverter defibrillator (AICD) implantation are one of the options for arrhythmia treatment. Right ventricular (RV) perforation is one of           the complications, which is rare. However, if it happens it can have serious consequences of PPM and AICD implantation.


We report a case of right ventricular perforation by an AICD lead in a 70-year-old Australian male underwent AICD implantation for non-ischemic cardiomyopathy for 18 days prior to the presentation.                The patient had right ventricular failure to capture. Computed tomography chest imaging and fluoroscopy demonstrated that the tip of the RV lead migrated outside the RV compared to previous study on implantation date. Hence RV perforation was diagnosed.  After catheter placement in pericardium was done for emergency event standby, the patient underwent RV defibrillator lead reposition from RV septum to RV apex without acute complication. Device interrogation after immediate re-operation and 3 years follow-up revealed a stable right ventricular pacing, threshold and position. 

 
     
     
     
     
 
    How to Cite  
     
  Kittichamroen, N. (2024). Defibrillator lead migration with subacute right ventricular perforation : A case report. Journal of Charoenkrung Pracharak Hospital20(2), 80–96. Retrieved from https://he02.tci-thaijo.org/index.php/JCP/article/view/269501  
     

 
 
     
     
 
    Issue  
     
  Vol. 20 No.2 (2567): July - December  

 
 
     
     
 
    Section  
     
  Case Report  

 
 
     
     
     

References

 
 

Piekarz J, Lelakowski J, Rydlewska A, Majewski J. Heart perforation in patients with permanent cardiac pacing - pilot personal observations. Arch Med Sci 2012;8(1):70-4. doi: 10.5114/aoms.2012.27284. PMID: 22457678; PMCID: PMC3309440.

Laborderie J, Barandon L, Ploux S, Deplagne A, Mokrani B, Reuter S, et al. Management of subacute and delayed right ventricular perforation with a pacing or an implantable cardioverter-defibrillator lead. Am J Cardiol 2008;102(10):1352-5. doi: 10.1016/j.amjcard.2008.07.025. Epub 2008 Sep 12. PMID: 18993154.

Khan MN, Joseph G, Khaykin Y, Ziada KM, Wilkoff BL. Delayed lead perforation: a disturbing trend. Pacing Clin Electrophysiol 2005;28(3):251-3. doi: 10.1111/j.1540-8159.2005.40003.x. PMID: 15733190.

Hirschl DA, Jain VR, Spindola-Franco H, Gross JN, Haramati LB. Prevalence and characterization of asymptomatic pacemaker and ICD lead perforation on CT. Pacing Clin Electrophysiol 2007;30(1):28-32. doi: 10.1111/j.1540-8159.2007.00575.x. PMID: 17241311.

Sivakumaran S, Irwin ME, Gulamhusein SS, Senaratne MP. Post pacemaker implant pericarditis: incidence and outcomes with active-fixation leads. Pacing Clin Electrophysiol 2002;25(5):833-7. doi: 10.1046/j.1460-9592.2002.t01-1-00833.x. PMID: 12049377.

Carlson MD, Freedman RA, Levine PA. Lead perforation: incidence in registries. Pacing Clin Electrophysiol 2008;31(1):13-5. doi: 10.1111/j.1540-8159.2007.00943.x. PMID: 18181903.

Rydlewska A, Małecka B, Zabek A, Klimeczek P, Lelakowski J, Pasowicz M, et al. Delayed perforation of the right ventricle as a complication of permanent cardiac pacing-is following the guidelines always the right choice? non-standard treatment-a case report and literature review. Kardiol Pol 2010;68(3): 357-61. PMID: 20411465.

Madershahian N, Wippermann J, Wahlers T. The bite of the lead: multiorgan perforation by an active-fixation permanent pacemaker lead. Interact Cardiovasc Thorac Surg 2010 ;11(1):93-4. doi: 10.1510/icvts.2010.234542. Epub 2010 Apr 13. PMID: 20388699.

Merla R, Reddy NK, Kunapuli S, Schwarz E, Vitarelli A, Rosanio S. Late right ventricular perforation and hemothorax after transvenous defibrillator lead implantation. Am J Med Sci 2007;334(3):209-11. doi: 10.1097/MAJ.0b013e31814254c5. PMID: 17873536.

Greenberg S, Lawton J, Chen J. Images in cardiovascular medicine. Right ventricular lead perforation presenting as left chest wall muscle stimulation. Circulation 2005;111(25):e451-2. doi: 10.1161/CIRCULATIONAHA.104.494732. PMID: 15983254.

Celik T, Kose S, Bugan B, Iyisoy A, Akgun V, Cingoz F. Hiccup as a result of late lead perforation: report of two cases and review of the literature. Europace 2009;11(7):963-5. doi: 10.1093/europace/eup071. Epub 2009 Apr 9. PMID: 19359331.

Laborderie J, Bordachar P, Reuter S, Clémenty J. Myocardial pacing lead perforation revealed by mammary hematoma next to the device pocket. J Cardiovasc Electrophysiol 2007;18(3):338. doi: 10.1111/j.1540-8167.2007.00724.x. PMID: 17319003.

Singhal S, Cooper JM, Cheung AT, Acker MA. Images in cardiovascular medicine. Rib perforation from a right ventricular pacemaker lead. Circulation 2007;115(14):e391-2. doi: 10.1161/CIRCULATIONAHA.106.669630. PMID: 17420357.

Epstein AE, Baker JH 2nd, Beau SL, Deering TF, Greenberg SM, Goldman DS. Performance of the St. Jude Medical Riata leads. Heart Rhythm 2009;6(2):204-9. doi: 10.1016/j.hrthm.2008.10.030. Epub 2008 Oct 25. PMID: 19103512.

Sterlinski M, Przybylski A, Maciag A, Syska P, Pytkowski M, Lewandowski M, et al. Subacute cardiac perforations associated with active fixation leads. Europace 2009;11(2):206-12. doi: 10.1093/europace/eun363. PMID: 19109359.

Vlay SC. Complications of active-fixation electrodes. Pacing Clin Electrophysiol 2002 ;25(8):1153-4. doi: 10.1046/j.1460-9592.2002.01153.x. PMID: 12358162.

Maziarz A, Ząbek A, Małecka B, Kutarski A, Lelakowski J. Cardiac chambers perforation by pacemaker and cardioverter-defibrillator leads. Own experience in diagnosis, treatment and preventive methods. Kardiol Pol 2012;70(5):508-10. PMID: 22623248.

Vlay SC. Concerns about the Riata ST (St. Jude Medical) ICD lead. Pacing Clin Electrophysiol 2008;31(1):1-2. doi: 10.1111/j.1540-8159.2007.00916.x. PMID: 18181899.

Danik SB, Mansour M, Heist EK, Ellinor P, Milan D, Singh J, et al. Timing of delayed perforation with the St. Jude Riata lead: a single-center experience and a review of the literature. Heart Rhythm 2008;5(12):1667-72. doi: 10.1016/j.hrthm.2008.09.017. Epub 2008 Sep 16. PMID: 19084802.

Lloyd MS, Shaik MN, Riley M, Langberg JJ. More late perforations with the Riata defibrillator lead from a high-volume center: an update on the numbers. Pacing Clin Electrophysiol 2008;31(6):784-5. doi: 10.1111/j.1540-8159.2008.01089_1.x. PMID: 18507558.

Hermanides R, Ottervanger J, Elvan A, Ramdat Misier AR. Life-threatening perforation of a defibrillation lead. Neth Heart J 2009;17(3):113-4. doi: 10.1007/BF03086230. PMID: 19325904.

Rordorf R, Canevese F, Vicentini A, Petracci B, Savastano S, Sanzo A, et al. Delayed ICD lead cardiac perforation: comparison of small versus standard-diameter leads implanted in a single center. Pacing Clin Electrophysiol 2011;34(4):475-83. doi: 10.1111/j.1540-8159.2010.03002.x. Epub 2011 Jan 5. PMID: 21208240.

Mahapatra S, Bybee KA, Bunch TJ, Espinosa RE, Sinak LJ, McGoon MD, et al. Incidence and predictors of cardiac perforation after permanent pacemaker placement. Heart Rhythm 2005;2(9):907-11. doi: 10.1016/j.hrthm.2005.06.011. PMID: 16171740.

Akyol A, Aydin A, Erdinler I, Oguz E. Late perforation of the heart, pericardium, and diaphragm by an active-fixation ventricular lead. Pacing Clin Electrophysiol 2005;28(4):350-1. doi: 10.1111/j.1540-8159.2005.09374.x. PMID: 15826277.

Lau EW, Shannon HJ, McKavanagh P. Delayed cardiac perforation by defibrillator lead placed in the right ventricular outflow tract resulting in massive pericardial effusion. Pacing Clin Electrophysiol 2008;31(12):1646-9. doi: 10.1111/j.1540-8159.2008.01240.x. PMID: 19067821.

Sassone B, Gabrieli L, Boggian G, Pilato E. Management of traumatic implantable cardioverter defibrillator lead perforation of the right ventricle after car accident: a case report. Europace 2009;11(7):961-2. doi: 10.1093/europace/eup111. Epub 2009 May 17. PMID: 19451098.

Stefanidis AS, Margos PN, Kotsakis AA, Papasteriadis EG. Three-dimensional echocardiographic documentation of pacemaker lead perforation presenting as acute pericarditis. Hell J Cardiol 2009;50(4):335-7. PMID: 19622505.

Ohlow MA, Lauer B, Brunelli M, Geller JC. Incidence and predictors of pericardial effusion after permanent heart rhythm device implantation: prospective evaluation of 968 consecutive patients. Circ J 2013;77(4):975-81. doi: 10.1253/circj.cj-12-0707. Epub 2012 Dec

PMID: 23269085.

Kirchgesner T, Ghaye B, Marchandise S, Le Polain de Waroux JB, Coche E. Iatrogenic cardiac perforation due to pacing lead displacement: imaging findings. Diagn Interv Imaging 2016;97(2):233-8. doi: 10.1016/j.diii.2015.03.011. Epub 2015 May 27. PMID: 26025159.

Balabanoff C, Gaffney CE, Ghersin E, Okamoto Y, Carrillo R, Fishman JE. Radiographic and electrocardiography-gated non contrast cardiac CT assessment of lead perforation: modality comparison and interobserver agreement. J Cardiovasc Comput Tomogr 2014 ;8(5):384-90. doi: 10.1016/j.jcct.2014.08.004. Epub 2014 Aug 23. PMID: 25301044.

Henrikson CA, Leng CT, Yuh DD, Brinker JA. Computed tomography to assess possible cardiac lead perforation. Pacing Clin Electrophysiol 2006;29(5):509-11. doi: 10.1111/j.1540-8159.2006.00385.x. PMID: 16689847.

Nazarian S, Beinart R, Halperin HR. Magnetic resonance imaging and implantable devices. Circ Arrhythm Electrophysiol 2013;6(2):419-28. doi: 10.1161/CIRCEP.113.000116. PMID: 23592868.

Banaszewski M, Stępińska J. Right heart perforation by pacemaker leads. Arch Med Sci 2012;8(1):11-3. doi: 10.5114/aoms.2012.27273. PMID: 22457667; PMCID: PMC3309429.